| Model Number |
1882040 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the blade found the green seal in the inner hub was in place and in good condition.The inner tip was broken off.The tip was not returned; however, the portion that became detached would have measured approximately 0.07¿ long x 0.06¿ diameter.There was minor wear on the inner shaft 1.00¿ from the distal face of the inner hub, which corresponds to the junction between the outer tube and front hub.There was deformation of the outer tube cutting edge; however, there was a radial indent at the distal most end.There were scratches on the returned stylet.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A healthcare professional (hcp) reported via a manufacturer representative that the blade was not suctioning during the operation.T he blade was used normally and a cleaner brush was used to unclog the tube.There was no patient impact.
|
| |
|