Model Number 1882040
Medical Device Problem Codes Use of Device Problem (1670); Obstruction of Flow (2423)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the blade found the green seal in the inner hub was in place and in good condition.The inner tip was broken off.The tip was not returned; however, the portion that became detached would have measured approximately 0.07¿ long x 0.06¿ diameter.There was minor wear on the inner shaft 1.00¿ from the distal face of the inner hub, which corresponds to the junction between the outer tube and front hub.There was deformation of the outer tube cutting edge; however, there was a radial indent at the distal most end.There were scratches on the returned stylet.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional (hcp) reported via a manufacturer representative that the blade was not suctioning during the operation.T he blade was used normally and a cleaner brush was used to unclog the tube.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7899505
Report Number1045254-2018-00452
Device Sequence Number2052250
Product Code EQJ
UDI-Device Identifier00681490613286
UDI-Public00681490613286
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1882040
Device Catalogue Number1882040
Device Lot Number0214921934
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2018
Initial Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-