| Model Number |
CIE3-11-24-00 |
| Medical Device Problem Codes |
Loss of Osseointegration (2408); Migration (4003)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
08/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the information known to date, the use of bmp during the index surgery caused an osteolytic response in the patient at the levels where the varilift-lx devices were implanted.The osteolytic response led to a migration of the devices and an additional surgery for removal of the devices 2.5 months post-op.There is no information to date to suggest that the a defect or malfunction of the varilift-lx caused or contributed to this event.The use of bmp with the varilift-lx devices is considered off-label use.
|
| |
|
Event or Problem Description
|
|
Varilift-lx removal 2.5 months post-op due to device migration caused by osteolytic response from bmp use at index surgery.
|
| |
|