We have received the device for evaluation and we were able to confirm the reported incident.We found a leak at the hub joint.This defect would not cause or contribute to a death or serious injury if the malfunction was to recur.However, during our evaluation, we observed the balloon in this catheter inflated partially but was unable to deflate when the plunger of the syringe was pulled.So, we are providing this report as this malfunction can cause or contribute to a serious injury if the issue was to recur had it been used in a patient.We found a glue like particle inside the catheter that prevented the saline from exiting the balloon.As a result, the balloon could not deflate.The patient's vascular rate would never be exposed to this particle since it was contained inside the lumen of the catheter.We could not conclusively determine the source of this particle.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.Further, we have not received any other complaints of similar nature for devices from this lot.Therefore, we believe that it was an isolated incident.Device was not used in the patient.
|