• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY Back to Search Results
Catalog Number PNML6F088804
Medical Device Problem Codes Device Damaged Prior to Use (2284); Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the neuron max had an unknown substance on the outside of the catheter shaft.The neuron max was ovalized approximately 66.5 and 81.5 cm from the hub.Conclusions: evaluation of the returned neuron max revealed that the sheath shaft had an unknown substance on it.The unknown substance could not be determined.Also, the hydrophilic coating was present on the catheter shaft.Further evaluation of the returned neuron max revealed that the shaft was ovalized.The ovalization was likely incidental to the reported issue and may have occurred while packaging the device for return to penumbra.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure, the physician noticed a "glue like substance" mid way up the shaft the neuron max 6f 088 long sheath (neuron max).The substance was found prior to use and, therefore, the neuron max was not used in the procedure.The procedure was completed using a new neuron max.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7900296
Report Number3005168196-2018-01865
Device Sequence Number169862
Product Code DQY
UDI-Device Identifier00814548012216
UDI-Public00814548012216
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/24/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2021
Device Catalogue NumberPNML6F088804
Device Lot NumberF81464
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2018
Initial Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-