• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAH
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call the insulin pump alarmed motor error and high blood glucose event.The customer¿s blood glucose level was unknown at the time of the incident.The customer stated that the drive support cap was normal and that the insulin pump was not exposed to high magnetic fields and was able to complete the rewind process.Displacement test was performed and passed.Customer also reported a high blood glucose of 424 mg/dl and no form of treatment was reported.No high blood glucose troubleshooting was performed.The customer was advised to monitor and call back for further assistance.The insulin pump is not expected to return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7900959
Report Number2032227-2018-32209
Device Sequence Number19220487
Product Code OYC
UDI-Device Identifier00643169476578
UDI-Public(01)00643169476578
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAH
Device Catalogue NumberMMT-723RNAH
Device Lot NumberA0723RNAHJ
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2018
Initial Report FDA Received Date09/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
Patient Weight185
-
-