WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-1-FEM-TULIP |
Device Problems
Unintended Movement (3026); Insufficient Information (3190); Appropriate Term/Code Not Available (3191); Migration (4003)
|
Patient Problems
Pulmonary Embolism (1498); Dyspnea (1816); Myocardial Infarction (1969); Internal Organ Perforation (1987); Pain (1994); Thrombosis (2100); Perforation of Vessels (2135); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog # is unknown but referred to as cook gunther tulip filter.Occupation: non-healthcare professional.(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a gunther tulip filter on (b)(6) 2012".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2012 via the right common femoral vein due to contraindication to anticoagulation and a risk of deep vein thrombosis post trauma.Patient alleges migration, tilt, organ perforation, pulmonary embolism.Patient further alleges heaving breathing, shortness of breathe, weakness, disorientation, heart attack caused by two large blood clots.
|
|
Manufacturer Narrative
|
Additional information: event, relevant tests/laboratory data, other relevant history.Exemption number: e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Patient code: pulmonary embolism (1498)- listed in the ifu, organ(s), perforation of (1987)- not listed in ifu, dyspnea (1816)-not listed in ifu.Device code: appropriate term/code not available (3191)- device perforation-not listed in the ifu, migration (4003)- not listed in ifu, unintended movement (3026)- listed in ifu.Mfr site: name and address for importer site: (b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information provided at this time.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer blank fields on this form indicate the information is unknown or unavailable, or unchanged.Mfr site: name and address for importer site: (b)(6).Registration no.: (b)(4).Additional information- investigation- it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tulip - mig, tilt, vc & organ perforation, pe, shortness of breath, heart attack, disorientation'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Manipulation in the area of the filter implant may cause migration or contribute to changes in the filter configuration and placement.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Pe is a known risk in relation to filter implant reported in the published scientific literature.Also, it is reported that the pulmonary embolism in some cases may originate from upper extremities instead of lower extremity veins.With all filters, there is some risk of further pulmonary embolism.Unknown if the reported shortness of breath is directly related to the filter.Unknown if the reported disorientation or heart attack is directly related to the filter and unable to identify corresponding failure mode(s) at this time.Lot# is unknown, but the device (igtcfs-65-1-fem-tulip) and tulip filter are manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
H6 device code(s): appropriate term/code not available (3191) was selected for the alleged device perforation and tilt.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date: migration, blood clots, pain, physical limitations/weakness, depression.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Unknown if the reported blood clots, pain, physical limitations/weakness, depression are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Lot number is unknown; however, there is no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Additional information received.The patient alleges that the filter was successfully removed on (b)(6) 2019 at the patient's request.Patient is alleging pulmonary embolism, vena cava perforation, tilt, and blood clots.The patient is also alleging that they cannot walk or stand for long periods of time.Need help walking stairs.Running or strenuous activity is an impossibility.Depression.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Corrected data: b5 and h6.Investigation: investigation is reopened due to additional information provided.Investigation is reopened due to additional information provided: thrombosis.The reported allegations have been further investigated based on the information provided to date.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Lot number is unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Heart attack caused by two large bloodclots.
|
|
Search Alerts/Recalls
|
|
|