• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDA-006
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/03/2018
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2018, a 6mm amplatzer duct occluder was selected for implant in an infant with a 6fr amplatzer torqvue delivery system.During device preparation, there was no issue loading the device.During placement, the user was unable to re-sheath the device as the occluder was caught on the sheath.Due to the failure to re-capture the device, the patient was sent for surgical removal of the device where the sheath was exchanged for a cook sheath and the device was pulled to the ivc and retrieved through the cook sheath.The case was aborted without closure of the defect.The delivery sheath was inspected and observed to be invaginated at the terminal end.
 
Additional Manufacturer Narrative
An event of an inability to retract the occluder into the delivery sheath was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key7901926
Report Number2135147-2018-00158
Device Sequence Number1552466
Product Code MAE
UDI-Device Identifier00811806011134
UDI-Public00811806011134
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number9-PDA-006
Device Catalogue Number9-PDA-006
Device Lot Number6292916
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/03/2018
Supplement Date Received by Manufacturer11/05/2018
Initial Report FDA Received Date09/24/2018
Supplement Report FDA Received Date11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age1 YR
Patient Weight10
-
-