• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 18IN 2-0 S/A PS-2 PRM; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 18IN 2-0 S/A PS-2 PRM; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 593H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2018 and suture was used.During the procedure, the suture was breaking.A different type device was used to complete the procedure with no adverse patient consequences.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).Date sent to the fda: (b)(4) 2018.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Additional information was requested and the following was obtained: (b)(4) lot lah125? - no suture was returned for code 585 for evaluation.2-was there an issue with the product (b)(4) lot kmp968 suture experienced breakage? - yes,sorry.From what i understand, the actual suture was breaking.Not the thread.3- please confirm which of the products should be considered for this complaint case.- i¿m assuming (b)(4) lot kmp968.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 18IN 2-0 S/A PS-2 PRM
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7902656
Report Number2210968-2018-75948
Device Sequence Number18874146
Product Code GAR
UDI-Device Identifier10705031006539
UDI-Public10705031006539
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number593H
Device Lot NumberKMP968
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/30/2018
Supplement Date Received by Manufacturer10/01/2018
Initial Report FDA Received Date09/24/2018
Supplement Report FDA Received Date10/11/2018
Patient Sequence Number1
-
-