Catalog Number 400SMTSTD0620
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, the physician felt resistance and was unable to advance a smart coil out of its introducer sheath, despite gently pushing and pulling the pusher assembly.The smart coil was therefore removed, and the procedure was completed using additional smart coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was undamaged.The introducer sheath was distal to the midjoint on the pusher assembly.The embolization coil was intact with the pusher assembly.The embolization coil was undamaged.Conclusion: evaluation of the returned smart coil revealed the pusher assembly mid-joint was retracted through the introducer sheath friction lock.The inner diameter of the introducer sheath friction lock is smaller than the rest of the introducer sheath and allows it to properly seat on the pusher assembly mid-joint.If the mid-joint is retract through the friction lock, resistance will likely be encountered while advancing the device out of the introducer sheath.This resistance may have contributed to the reported inability to advance the device out of its introducer sheath.During functional testing, the smart coil was able to advance out of its introducer sheath and through a demonstration microcatheter without issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7902723
Report Number3005168196-2018-01890
Device Sequence Number1434524
Product Code HCG
UDI-Device Identifier00814548014678
UDI-Public00814548014678
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/19/2022
Device Catalogue Number400SMTSTD0620
Device Lot NumberF79583
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer01/15/2019
Initial Report FDA Received Date09/24/2018
Supplement Report FDA Received Date02/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-