Model Number 18760
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, there was an air leaking when inflating the device¿s cuff.It was reported that the tip part of the cuff seemed to be distorted when inflated.The customer reported that there was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one sample of device was received for evaluation.It was observed the cuff deflated immediately.A visual inspection was conducted and it was observed the cuff presents a cut.The reported event was confirmed.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLINCKRODT
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key7903100
Report Number2936999-2018-00607
Device Sequence Number1394985
Product Code BTR
UDI-Device Identifier80884522007572
UDI-Public80884522007572
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2023
Device Model Number18760
Device Catalogue Number18760
Device Lot Number18A0449JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Initial Date Received by Manufacturer 09/03/2018
Supplement Date Received by Manufacturer10/26/2018
Initial Report FDA Received Date09/24/2018
Supplement Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-