AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012631-39
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a cause for the reported stent dislodgment.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat an unspecified moderately calcified peripheral artery.An 8.0 x 39 mm omnilink elite vascular balloon-expandable stent system was used and did not meet any resistance during advancement.However, the stent dislocated from the stent system; therefore, it was implanted in healthy tissue.Another unspecified omnilink elite stent system was then used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7903107
Report Number2024168-2018-07344
Device Sequence Number1551964
Product Code NIO
UDI-Device Identifier08717648178511
UDI-Public08717648178511
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number1012631-39
Device Lot Number8042341
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2018
Initial Report FDA Received Date09/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-