MAUDE Adverse Event Report: BARD ACCESS SYSTEMS BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE 18F X 1.7CM; TUBE, GASTRO-ENTEROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE 18F X 1.7CM; TUBE, GASTRO-ENTEROSTOMY Back to Search Results
Model Number 000282
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during the placement procedure the button allegedly leaked.A new button was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was performed.The lot met all release criteria.Investigation summary: based on the sample evaluation: one bard button replacement gastrostomy tube kit ,18fr 1.7 cm was returned for evaluation.Visual and functional evaluations were performed.The investigation is inconclusive for button valve leak, as the button device was attached to the continuous feeding adaptor found with the complaint sample and flushed with a syringe filled with water.No leaks occurred in the assembly.The assembly was then aspirated by pulling on the syringe handle.Water was held throughout the assembly and no obvious leaks with the valve were found.The button dome was removed.No visual defects were noticed.However, the clinical conditions are unknown which could have caused the button valve to leak, therefore the complaint is considered inconclusive.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review:a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during the placement procedure the button allegedly leaked.A new button was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The investigation is currently under way.
 
Event or Problem Description
It was reported that during the placement procedure the button allegedly leaked.A new button was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE 18F X 1.7CM
Common Device Name
TUBE, GASTRO-ENTEROSTOMY
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7903787
Report Number3006260740-2018-02561
Device Sequence Number11189421
Product Code KGC
UDI-Device Identifier10801741087858
UDI-Public(01)10801741087858
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K904779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000282
Device Catalogue Number000282
Device Lot NumberHUBS1188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer11/29/2018
03/29/2019
Initial Report FDA Received Date09/24/2018
Supplement Report FDA Received Date12/14/2018
04/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age18 MO
-
-