A customer in the united states notified biomérieux of discrepant susceptibility results for cap survey strain (d-08) when testing with the vitek® 2 ast-gn70 test kit.The expected result for this ungraded survey alcaligenes faecalis strain is susceptible.On initial testing, the strain tested false resistant (mic >=128) for piperacillin/tazobactam using vitek 2 ast-gn70, lot 5900678403.As there is no patient associated with this quality control strain, there is no adverse event related to any patient's state of health.A biomérieux internal investigation has been initiated.
|
|
A customer in the us notified biomérieux of false resistant results for cap survey strain (d-08) when testing with the vitek 2 ast-gn70 cards.The alcaligenes faecalis ssp.Faecalis strain resulted in mics => 128 (r), <= 4 (s), and 64 (i) for piperacillin/tazobactam (tzp) with their original strain and 64 (i), <= 4 (s) with their back-up swab.The expected result was susceptible.The strains were submitted for investigation.Investigation details: the customer's submitted strains and the biomerieux internal cap d-08 survey strains were subcultured on tsab agar, identifications confirmed as a.Faecalis ssp faecalis, and testing included ast-gn70 cards from customer lots (5900755403, 5900678403) and a random lot (5900623403) as well as broth microdilution (bmd) as the reference method for tzp03n formulation on this card.911733 (customer's original d-08 strain): two cards tested gave a resistant mic => 128 and the third card mic = 64 (i).Bmd gave a susceptible mic <= 4.911734 (customer's back-up d-08 strain): all three cards tested gave a resistant mic => 128.Bmd gave a susceptible mic <= 4.911704 (biomerieux internal d-08 strain): on two of the ast-gn70 cards, tzp mic = 64 (i) and on the third card mic <= 4 were obtained.Bmd gave a susceptible mic <= 4.Card and reference method are in agreement for one card while demonstrating an essential agreement error with a minor category error for the remaining two cards.Qs performed a tw search for all complaints against ast-gn70 card lots 5900755403 and 5900678403 returning three and one complaint respectively.A review of these complaints did not indicate a trend for these card lots.This investigation duplicated the customer's inconsistent results.This atypical survey strain will be added to the r&d stock collection.
|