| Model Number |
CRPLUS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer was provided with a replacement device.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had all three icons (charge pak, attention and service wrench) present on its readiness display.The three icons being illuminated indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had all three icons (charge pak, attention and service wrench) present on its readiness display.The three icons being illuminated indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio determined the cause of the reported issue was due to a shorted capacitor, designated c76 on the digital pcb assembly.
|
| |
|