The reported event of an inability to advance the occluder through the delivery system could not be confirmed.The nitinol braiding at the proximal end of the occluder was bunched, which contributed to difficulty loading the occluder into the delivery sheath.As this damage is consistent with a forceful attempt to advance the occluder, and potentially contributed to the advancement difficulty, it is not possible to determine the extent of this difficulty or the bunching of the nitinol wires prior to use in the field.Additionally, despite this damage, the occluder was able to be successfully deployed through a stock 6f delivery system, meeting functional specifications.The occluder met dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event could not be conclusively determined.
|