| Model Number |
MMT-7020A |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 132 mg/dl and the sensor glucose was 88 mg/dl.Insulin delivery was suspended due to sensor glucose values.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened and used sensor and performed continuity resistance test.Sensor passed per specification.Also, performed bicarbonate buffer test.Sensor passed per specifications with accurate readings.See attached file for details.
|
| |
|