| Catalog Number |
AUBS-4-6 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
It was initially reported via an user facility 3500a report received on 12sep2018, an ascend aq - balloon lost its integrity.A new balloon had to be opened to complete the procedure.Product had patient contact, but no patient harm.Product is available to be returned to the manufacturer.Additional information after follow up with the user facility was received on 17sep2018 via email.The intended procedure was a right ureteroscopy with holmium laser ablation of stone, retrieval of stone fragments, and placement of ureteral stent.No patient harm was reported.
|
| |
|
Event or Problem Description
|
|
There has been no new information received since the last report was submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection and functional testing of the returned device was conducted.A document based investigation was also performed including a review of complaint history, the device history record, instructions for use, quality control data, and specifications.One ascend® aq® ureteral balloon catheter was returned to manufacturer for investigation.The returned packaging confirms the complaint device lot number 7570408.The device was returned with distal tip inserted into the hub.Functional testing was performed with water inserted into the balloon using an inflation device.The balloon material was determined to have a pin hole leak.Visual exam notes the leak in the balloon material is around the distal metal band.A review of the device history record found there were no non-conformances related to the reported failure.A review of complaint history records revealed this is the only complaint associated with the complaint device lot number 7570408.A functional test on the aubs-4-6 revealed a pinhole leak in the balloon.A visual examination noted the pinhole leak in the balloon material is near the distal metal band.The cause of the pinhole leak in the balloon material is unknown however it is possible the device was punctured during the procedure.There is no indication that a design process or related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Per the quality engineering risk assessment, no further action is warranted.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|