Model Number F5
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation revealed that the root cause of this failure mode is contributed to "component failure".While the patient was in the wheelchair the left side drive wheel detached from the device.The patient was not injured in the event.The wheelchair was evaluated and the main hub bolt loosened and backed out, allowing the drive wheel to slide off the drive motor shaft.This is a known issue that has since been corrected.The failure was linked back to the supplier and related to an assembly error installing the wheel hub to the drive motor shaft.The risk for this failure to occur is low and the likely hood of injury to occur is also very low.Due to this known repeat failure mode, a capa has been opened by our parent company.Upon completion of the capa, if further corrective action is required for this wheelchair this case will be re-opened and a follow-up mdr will be submitted.The spare parts necessary for the repair have been supplied and the wheelchair repaired and returned to the patient.The dhr for this device has been reviewed and the wheelchair met specification prior to distribution.
 
Event or Problem Description
Reports that the left drive wheel assembly detached from the wheelchair.It was discovered that the main hub bolt had backed out.No injuries have occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERMOBIL F5
Common Device Name
POWERED WHEELCHAIR
Manufacturer (Section D)
PERMOBIL INC.
300 duke drive
lebanon TN 37090
MDR Report Key7905143
Report Number1221084-2018-00057
Device Sequence Number13905041
Product Code ITI
UDI-Device Identifier17330818229844
UDI-Public17330818229844
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K143014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberF5
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2018
Initial Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight68
-
-