| Model Number |
265720 |
| Medical Device Problem Code |
Material Invagination (1336)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the catheter mushroomed while being injected with water.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the catheter mushroomed while being injected with water.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use was found adequate and states the following: "warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex ad cause balloon to burst.Do not exceed recommendation capacities.".Section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|