|
Reportable based on device analysis completed on 04-sep-2018.It was reported that tip damaged occurred.The stenosed target lesion was located in posterior tibial artery.The target lesion was delivered thru a v-18 wire and a non bsc sheath.A 6.0mmx100mmx135cm-hybrid fg sterling otw was used to dilate the lesion.When balloons were attempted, a kink in the catheter near the tip of the balloon was noticed.The procedure was completed with a different device.No patient complications were reported.It was further reported that the returned device analysis revealed a hole in the inner shaft from the tip.Device evaluated by manufacturer: the balloon was loosely folded and there was blood on the balloon.Microscopic examination revealed a hole in the inner shaft 5.5cm from the tip, kinked balloon and inner shaft at 9.7cm from the tip.
|