• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHILON SUTURE 36IN(90CM) 3-0 BLK; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHILON SUTURE 36IN(90CM) 3-0 BLK; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Catalog Number F2415H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Additional information was requested and the following was obtained: did 2 sutures had the issue in 1 procedure? two units (sutures) are implicated for the this procedure.One unit is implicated for another/2nd procedure (captured separately).When did the needles pull off in the package/during dispensing/ during use on patient (suturing)? during use on animal when performing the cutaneous suture.Specific number of procedures in which the needles pulled off, during (b)(6) 2018.Two procedures during (b)(6).No further information available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported an animal underwent an unknown procedure on an unknown date and suture was used.Needles from the same box pulled off.No further information available.No patient consequences reported.
 
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).Date sent to the fda: (b)(4) 2018.Unopened samples of product code were returned for analysis.During the visual inspection of thirteen unopened samples, no defects were found on the packages.The samples were opened and the swage and attachment area of the needles were as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed and the manufacturing/ packaging criteria were met prior to the release of this batch.Per the conditions of the samples received, no attachment defects were found, and the tested samples met the finished goods requirements,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHILON SUTURE 36IN(90CM) 3-0 BLK
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7906075
Report Number2210968-2018-75985
Device Sequence Number1692987
Product Code GAR
UDI-Device Identifier10705031238930
UDI-Public10705031238930
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Catalogue NumberF2415H
Device Lot NumberLEH029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2018
Initial Date Received by Manufacturer 08/30/2018
Supplement Date Received by Manufacturer10/01/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date10/23/2018
Patient Sequence Number1
-
-