| Model Number |
CC4204A +21.0 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon noticed a cc4204a, +21.0 diopter, intraocular lens had a tear during insertion.It was reported that there was patient contact but no patient injury.There is no certainty to the cause of the event.Backup lens was inserted without any reported problems.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: device evaluation: lens was returned in liquid, in lens vial.Visual inspection found the haptic torn.The nanopoint was returned in a biohazard tag.Visual inspection found the plunger bent and no visible damage to cartridge.(b)(4).
|
| |
|