• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - CARTAGO CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CARTAGO CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2H8537
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an unknown quantity of access-continu-flo sets would not flow when attempting to prime the tubing.The user inverted the filter and tapped all check valves.It would not flow even after the spike was pushed further, and saline bag and chamber squeezed repeatedly.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The device was received for evaluation.The tubing of the second y-site is separated from the y-site.During sample analysis, excess solvent was observed in the tube.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
MDR Report Key7906187
Report Number1416980-2018-06130
Device Sequence Number13873791
Product Code FPA
UDI-Device Identifier00085412080901
UDI-Public(01)00085412080901
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K961225
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/12/2023
Device Catalogue Number2H8537
Device Lot NumberR18D12080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Initial Date Received by Manufacturer 08/31/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date11/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-