| Catalog Number |
VBHR080702A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
08/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
|
| |
|
Event or Problem Description
|
|
The following was reported to gore: the patient presented for a thoracic case.The gore® viabahn® endoprosthesis with heparin was placed in the left exterior iliac to the common iliac to be a conduit for the advancement of the cook device.The doctor reported the vessel was very tortuous.Next a cook alpha 34-113 (18fr-7.1mm) device was advanced to the gore® viabahn® endoprosthesis with heparin.Attempts to advance the cook device through the gore® viabahn® endoprosthesis with heparin were unsuccessful.During advancement it appeared that the gore® viabahn® endoprosthesis with heparin 'accordioned'.The cook device was withdrawn and the doctor observed that the gore® viabahn® endoprosthesis with heparin was attached to it.The absence of the gore® viabahn® endoprosthesis with heparin created an iliac rupture.Control of the rupture was achieved with a coda balloon.The iliac was repaired using multiple endovascular devices and the thoracic procedure was completed.The doctor reported to gore that the patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: review of the manufacturing records indicated the device met pre-release specifications.The device was not returned.Consequently, a direct product analysis was not possible.All information has been placed on file for use in tracking and trending.Method code 2, results code 2.Corrected data: result code 1.
|
| |
|
Search Alerts/Recalls
|