| Medical Device Problem Codes |
Partial Blockage (1065); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a occlusion (frequent/persistent) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 28-sep-2018 with the following findings: there were multiple occlusion alarms observed in the pump's black box with an associated high force.The prime steps were performed successfully with no alarms.A force sensor calibration test confirmed that the sensor was detecting the correct force.The pump was exercised for 24 hours with no occlusions observed.Unrelated to the initial allegation, the battery compartment was cracked and the display screen was dim and discolored.
|
| |
|