Model Number CQ2015A
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.The lens was returned dry, loose in the box.There was clear surgical residue on the product and lens surface.Visual inspection found the optic torn.No similar complaint type events reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that as the cq2015a intraocular lens, diopter +23.0 was being injected into the patient's left (os) eye, the haptic broke inside of the lens inserter.Reportedly, this was due to the lens being "misloaded by the tech." the lens was explanted intraoperatively.This occurred on (b)(6) 2018.
 
Additional Manufacturer Narrative
Additional data: the lens was removed and exchanged with a backup lens intraoperatively on (b)(6) 2018.Corrected data: return date is corrected to 23-aug-2018.Device evaluation: visual inspection found the optic torn along with the haptic torn.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEE H10
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7906884
Report Number2023826-2018-01327
Device Sequence Number19284388
Product Code HQL
UDI-Device Identifier00841542102049
UDI-Public00841542102049
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P990013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Model NumberCQ2015A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Initial Date Received by Manufacturer 08/10/2018
Supplement Date Received by Manufacturer08/10/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date10/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
-
-