As reported to angiodynamics on (b)(6) 2018: during an evlt procedure, when inserting the guidewire into the catheter, the guidewire unraveled.The device was set aside and a new of the same was used to complete the procedure.The patient was not harmed or injured due to this event.The disposable device has been returned to the manufacturer for a device evaluation.
|
|
Returned for evaluation was a.035"x150cm double ended guidewire.The device was forwarded to angiodynamics' supplier of this product for a device evaluation and root cause determination.Per the supplier's evaluation, "both the distal and proximal ends of the double ended guidewire appeared to have been shaped by human intervention, when the shape and the location of the bends were analyzed.A review of the dhr of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspected tested at lake region medical limited and was determined to be acceptable".The customer's reported complaint description of guidewire "frayed" is confirmed.Although the complaint description is confirmed, a definitive root cause cannot be determined.The most likely root cause is handling damage by the end user.A review of the angiodynamics' device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number contains the following statement: "using standard seldinger technique gain access to the desired vein with the needle and corresponding guidewire.If applicable, use the micro-access introducer to exchange for a larger diameter wire.Advance the 4fr trè-sheath introducer, over the wire, to the treatment location.If treating the great saphenous vein, position sheath tip so that it is 1-2cm below the sapheno-femoral junction.Verify sheath positioning using ultrasound guidance.Remove the dilator and guidewire." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference complaint #: (b)(4).
|