(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.Post-procedure photos of the implanted stent were returned and reviewed by an abbott clinical specialist.The reviewer concluded the following: the absolute pro stent image provided does appear to show a discontinuity within the stent pattern at roughly the middle section of the stent.Unfortunately, the imaging detail is not sufficient to confirm an actual separation of the stent cut pattern.Due to the heavy calcification in the vessel, there is likely some substantial influence on the stent pattern which could be what is appearing to be considered a fracture.Without additional higher detailed imaging, the confirmation of a fracture is unclear.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the complaint handling database revealed no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported stent fracture.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|