OBERDORF SYNTHES PRODUKTIONS GMBH SMALL HEXAGONAL SCREWDRIVER 2.5MM WIDTH ACROSS FLATS; SCREWDRIVERS Back to Search Results
Catalog Number 314.070
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information is unknown.Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes (b)(4) reports an event as follows: it was reported that on an unknown date, a hexagonal screwdriver with groove broke while being used in an unknown procedure.There was no patient consequence reported; it is unknown if there was a surgical delay or what the surgical/patient outcome were.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Additional narrative: corrected data: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part: 314.070 , lot: 9734993 , manufacturing site: hägendorf , release to warehouse date: 23 december 2015.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Only top level of the device history record reviewed as sub-components are not lot tracked.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Visual investigation: the screwdriver-hexagonal ¿2.5 w/grooved tip is completely broken off.The broken off portion is missing.The plastic handle shows wear of frequent use.Dimensional inspection: because of the existing use related damage on the available instrument and the missing broken off portion a dimensional inspection cannot be conducted and is not required per selected investigation flow.Drawing/specification review: the cause of the complained malfunction is a post-manufacturing caused breakage/damage due to mechanical overload to the device, therefore no drawing/specification review is required per selected investigation flow.Conclusion: the received condition agrees with the complaint description and the complaint therefore is confirmed.The more than eight years old multiple use device was subjected to frequent use.We are not able to determine an exact cause of breakage, but we suppose that at the time of surgery the device was subjected to mechanical overloading.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL HEXAGONAL SCREWDRIVER 2.5MM WIDTH ACROSS FLATS
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7907074
Report Number8030965-2018-56639
Device Sequence Number11176070
Product Code HXX
UDI-Device Identifier07611819019829
UDI-Public(01)07611819019829
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.070
Device Lot Number9734993
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2018
Initial Date Received by Manufacturer 09/03/2018
Supplement Date Received by Manufacturer09/13/2018
11/06/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date11/06/2018
11/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-