• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION SHOULDER; GLENOID, UNIFORM, PEGGED 42MM CM FOUNDATION SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION SHOULDER; GLENOID, UNIFORM, PEGGED 42MM CM FOUNDATION SHOULDER Back to Search Results
Model Number 500-04-142
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/30/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient's shoulder being infected.The surgeon removed the components and implanted cement spacers.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to an infection.The previous surgery and the revision detailed in this investigation occurred 4.5 years apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION SHOULDER
Common Device Name
GLENOID, UNIFORM, PEGGED 42MM CM FOUNDATION SHOULDER
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7907480
Report Number1644408-2018-00905
Device Sequence Number1585861
Product Code KWS
UDI-Device Identifier00888912143196
UDI-Public(01)00888912143196
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K960906
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2015
Device Model Number500-04-142
Device Catalogue Number500-04-142
Device Lot Number741C1004
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/30/2018
Supplement Date Received by Manufacturer10/12/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1309; 520-01-010, LOT 455C1106; 520-46-116, LOT 923C1051
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age74 YR
-
-