| Catalog Number |
1000071 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Debris found on single use instrument pack (b)(4) (outside and on instrument), which is likely to peel off from the red label.
|
| |
|
Event or Problem Description
|
|
Debris found on single use instrument pack 100-0071 (outside and on instrument), which is likely to peel off from the red label.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that star instruments kit sterile packaged was alleged of 'laser marking/printing fading/flaking' could be confirmed.Based on investigation, the root cause could have been caused by the contact of the steri-dot with "acetone", which is used as a cleaning agent during manufacturing.The device inspection revealed the following: the returned sterile inner pouches were returned and do indeed show some red debris all over them.It can be confirmed that the red debris are coming from the sterilization dots since the well-known red pattern of the gamma sign can be identified.
|
| |
|