Catalog Number M0036125150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that when the coil (subject device) was being pulled back for repositioning, the coil broke inside the microcatheter.The broken coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.The subject device is not available; therefore, functional testing as well as physical analysis cannot be performed.Information available indicated that the device was confirmed to be in good condition prior to use.Based on the information currently available, the exact cause for the reported event cannot be determined.
 
Event or Problem Description
It was reported that when the coil (subject device) was being pulled back for repositioning, the coil broke inside the microcatheter.The broken coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.Visual inspection was performed on the returned sample and it was observed that the main coil was seen to be protruding from the distal end of the catheter.The main coil was removed and noted to extremely stretched and broken at the proximal end.Functional testing could not be performed due to the damaged condition of the returned coil.It is likely that the main coil was damaged during use thus limiting the performance of the coil during the clinical procedure.Therefore, an assignable cause of component failure has been assigned to this investigation.
 
Event or Problem Description
It was reported that when the coil (subject device) was being pulled back for repositioning, the coil broke inside the microcatheter.The broken coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 5MM X 15CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7907623
Report Number3008881809-2018-00418
Device Sequence Number16512280
Product Code HCG
UDI-Device Identifier07613327128291
UDI-Public07613327128291
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2021
Device Catalogue NumberM0036125150
Device Lot Number20504117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer11/15/2018
04/15/2019
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date11/15/2018
04/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNKNOWN MANUFACTURER)
-
-