STRYKER ORTHOPAEDICS-MAHWAH 6.5 CANCELLOUS BONE SCREW 16MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 2030-6516-1
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Inflammation (1932); Injury (2348); Reaction (2414); Device Embedded In Tissue or Plaque (3165)
Date of Event 07/09/2008
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Photographs of explanted components show one intact screw and two screws with missing heads noted.Since it is unknown which screws were broken, all screws have been reported.Review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Conclusion: no readable post-operative x-ray of the primary right total hip arthroplasty prior to the (b)(6) 2008 post head fracture films to allow for evaluation of the initial acetabular position and seating of the alumina insert are available for review.In a ceramic/ceramic bearing a vertical acetabular position can contribute to mechanical failure.Delay in revision surgery from the time of the diagnosis of the ceramic bearing failure on (b)(6) 2008 until surgery on (b)(6) 2008 with the patient remaining active was largely responsible for the excessive fragmentation, metallosis, and diffuse distribution of the particular debris.There is nothing unique in the stryker products compared to competitive ceramic/ceramic bearings that contributed to the complications in this case which were increased by delay in revision surgery.Based upon the information available for review, no determination can be made regarding the cause of the initial fracture of the ceramic bearing four years after implantation in this case.The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.
 
Event or Problem Description
It was reported through the attorney for the patient, as a result of a lawsuit, that allegedly the patient received a trident ceramic on ceramic hip system.It was further alleged that, the patient began to experience some difficulties, requiring a revision surgery on (b)(6) 2008." update as per medical review: "on (b)(6) 2008 a revision of the right total hip was performed with removal of loose, fractured ceramic pieces.Extensive metallosis was noted and a revision total hip was performed for a diagnosis of fractured ceramic head and liner, displaced metal liner, erosion superior acetabular shell, and extensive wear debris.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5 CANCELLOUS BONE SCREW 16MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7907783
Report Number0002249697-2018-03028
Device Sequence Number13905034
Product Code HWC
UDI-Device Identifier07613327036978
UDI-Public07613327036978
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K894124
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2008
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2008
Device Catalogue Number2030-6516-1
Device Lot Number89137302
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2018
Initial Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age49 YR
Patient Weight76
-
-