Catalog Number 2H8537
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the spikes of an unknown quantity of clearlink sets were found loose (separated from the tubing) in the package.This was identified prior to use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The device was received for evaluation.Visual inspection on the returned sample revealed the tubing was separated from the male luer.Dimensional measurements were performed to the tubing and the results were within product specifications.Further examination of the tubing, using a scope, revealed that solvent wasn¿t applied uniformly.The cause of the condition was determined to be incorrect functioning of the machine solvent system.The reported condition was verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
MDR Report Key7907864
Report Number1416980-2018-06146
Device Sequence Number1625073
Product Code FPA
UDI-Device Identifier00085412080901
UDI-Public(01)00085412080901
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K961225
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/13/2023
Device Catalogue Number2H8537
Device Lot NumberR18B13090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Initial Date Received by Manufacturer 08/31/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date11/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-