| Catalog Number |
110135 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
During femur surface burring.Burr caught tip of surgeons glove while she pressed trigger.No harm to surgeon, just caught glove.Case type: pka.Picture is not from during surgery - it shows us trying to replicate what may have happened.It looked like her glove was loose and got caught up by the burr as power was engaged.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: mics, 6mm ball burr; during femur surface burring, burr caught tip of surgeons glove while she pressed trigger.Method & results: device evaluation and results: no device inspection could be completed as the product was not available for evaluation.Device history review: not performed as the mics, 6mm ball burr is an oem product.Complaint history review: based on the device identification, the catsweb and trackwise complaint databases were reviewed from 2011 to present for similar reported events regarding mics, 6mm ball burr; during femur surface burring, burr caught tip of surgeons glove while she pressed trigger failure of p/n: 110135, lot #: 17277027.There have been no other similar events for the referenced lot number.Conclusions: the failure mode could not be confirmed because the part was not available for evaluation.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: as the event did not involve a manufacturing related product problem indicating a non-conformity, adverse trend, or unanticipated hazard, no corrective action is required at this time.Device not returned.
|
| |
|
Event or Problem Description
|
|
During femur surface burring.Burr caught tip of surgeons glove while she pressed trigger.No harm to surgeon, just caught glove.Case type: pka.Picture is not from during surgery - it shows us trying to replicate what may have happened.It looked like her glove was loose and got caught up by the burr as power was engaged.
|
| |
|