• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT0203
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The embolization coil was intact with its pusher assembly.There was no functional allegation for the returned smart coil.Therefore, the smart coil was not functionally tested.Conclusions: evaluation of the returned smart coil revealed that two introducer sheaths were returned.These introducer sheaths were fully intact.The reported complaint of a fractured introducer sheath could not be confirmed.Further evaluation of the returned smart coil device history record revealed the number of devices and sheaths pulled for production reconciled correctly.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a coil embolization procedure, the physician found that the introducer sheath of a penumbra smart coil smart coil was broken into two pieces upon removal from the packaging hoop.The damage to the smart coil introducer sheath was found prior to use and, therefore, the smart coil was not used in the procedure.The procedure was completed using a new smart coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7908235
Report Number3005168196-2018-01902
Device Sequence Number19253210
Product Code HCG
UDI-Device Identifier00814548015279
UDI-Public00814548015279
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2021
Device Catalogue Number400SMTXSFT0203
Device Lot NumberF72738
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2018
Initial Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-