Model Number AR40M
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Date of Event 08/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.If implanted; give date: n/a (not applicable).The lens was removed/replaced within the initial procedure.If explanted; give date: n/a (not applicable).The lens was removed/replaced within the initial procedure.It was indicated that the device is not returning for evaluation; therefore a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Incision enlarged.Attempt has been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the trailing haptic of a model ar40m 4.5 diopter intraocular lens (iol) was not there once the lens was implanted into the patient's eye.The iol was removed and replaced with the backup lens during the same procedure.An incision enlargement was required.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).Additional info: based on receiving new information, the following fields have been updated accordingly.Age/date of birth: (b)(6).Sex/gender: female.Initial reporter: dr.(b)(6), health professional: yes.Occupation: physician.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7908349
Report Number2648035-2018-01258
Device Sequence Number1585874
Product Code HQL
UDI-Device Identifier05050474514089
UDI-Public(01)05050474514089(17)230621
Combination Product (Y/N)Y
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2023
Device Model NumberAR40M
Device Catalogue NumberAR40E00045
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/27/2018
Supplement Date Received by Manufacturer09/26/2018
10/25/2020
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date10/24/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE, EMERALDC30, LOT NUMBER UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Date Report Sent to FDA01/01/2005
-
-