Catalog Number UNK_LIM
Medical Device Problem Codes Break (1069); Fracture (1260); Loss of Osseointegration (2408)
Health Effect - Clinical Codes Injury (2348); Inadequate Osseointegration (2646)
Date of Event 08/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that patient's right knee was revised due to aseptic loosening of the femur and tibia, as well as a broken femoral stem at the stem/ offset bolt interface.Rep reported the patient's entire knee was revised.
 
Additional Manufacturer Narrative
Non bone interfacing device reported in error.Updated information for unknown bone interfacing device (femoral stem) provided corrections to: d1, d2, d4, d6, g5, h4 and h6 codes an event regarding femoral and tibial loosening as well as a fracture of the femoral stem involving an unknown stem component was reported.The event was confirmed based on clinician review of the x-rays provided.Method & results: -device evaluation and results: not performed as the device was not returned for evaluation.-medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: x-ray confirms fractured femoral stem; need revision operative report and examination of explanted components.-device history review: not performed as the lot id was not provided.-complaint history review: not performed as the lot id was not provided.Conclusions: review of the provided medical records by a clinical consultant stated the following comment: x-ray confirms fractured femoral stem.The exact cause of the event could not be determined because insufficient information was provided.Additional information, including revision operative report and examination of explanted components are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
It was reported that patient's right knee was revised due to aseptic loosening of the femur and tibia, as well as a broken femoral stem at the stem/ offset bolt interface.Rep reported the patient's entire knee was revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN FEMORAL STEM COMPONENT
Common Device Name
UNKNOWN KNEE IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7908433
Report Number0002249697-2018-03037
Device Sequence Number16564988
Product Code KRO
UDI-Device Identifier07613327045208
UDI-Public07613327045208
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K002552
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year1996
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Catalogue NumberUNK_LIM
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer10/04/2018
Initial Report FDA Received Date09/25/2018
Supplement Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
Patient Weight91
-
-