| Catalog Number |
48280232 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
08/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during tlif surgery, a short blade screw was inserted via k-wire without any problems.However, it was found by post operative radiograph that a part of the k-wire was being left in s1.The fragment of k-wire was not retrieved.No adverse consequences to the patient is reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed via email correspondence with the sales rep.The product was discarded and lot# not provided.It was reported that the awl was not used to prepare the pedicle.The k-wire guide tube was not used to insert the k-wire.The k-wire was inserted via j-probe outer tube the k-wire was properly held in position while removing the outer shaft of the needle.The k-wire was properly held in position while removing the dilator.It is unknown is a tip of the k-wire pinched during the insertion of an awl.The device surgical technique recommends that the awl is used to create a pilot hole prior to tapping.Conclusion: the most likely root cause of the reported event it is non-use of awl to prepare the pedicle.
|
| |
|
Event or Problem Description
|
|
It was reported that during tlif surgery, a short blade screw was inserted via k-wire without any problems.However, it was found by post operative radiograph that a part of the k-wire was being left in s1.The fragment of k-wire was not retrieved.No adverse consequences to the patient is reported.
|
| |
|
Search Alerts/Recalls
|