| Model Number |
MMT-523NAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced high blood glucose level of 426 mg/dl.Insulin pump had motor error alarm.Drive support cap was normal.Customer was able to rewind.Insulin pump had no delivery alarm and it was resolved by changing complete set.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No motor error alarm or unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.(b)(4).
|
| |
|