Evaluation summary: the silverhawk was returned for evaluation along with three detached cutter driver.Two cutter drivers had medtronic printed and the other had ev3 printed.No other ancillary devices were included.The silverhawk was inspected.The cutter was located back within the cutter window.No damage or anomalies to the silverhawk were noted.The distal assembly was inspected under microscope and a piece of metal protruding out from the cutter window, behind the cutter head, was identified.The piece of metal appeared to be a strut from a stent.The piece was attempted to be removed using a tweezers, however resistance was encountered.Direct light applied to the cutter housing observed the plunger was pushed to the proximal end of the housing against the distal rim of the cutter assembly.The silverhawk was connected to a returned cutter driver and activated.The cutter was able to be advanced into the housing.If information is provided in the future, a supplemental report will be issued.
|