Model Number 25-40-45 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).Livanova (b)(4) requested the device back for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that the values of a s5 gas blender system deviate during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
The device was returned to livanova (b)(4) for further investigation and repair.During the evaluation the reported issue could be confirmed and reproduced.As root cause a defective mass flow controller could be determined.The defective component was replaced to resolve the reported issue.Subsequent functional verification testing was completed without further issues.
|
|
Event Description
|
See initial.
|
|