• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 16FR DILATOR; DH EF PERC PLACEMENT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 16FR DILATOR; DH EF PERC PLACEMENT PRODUCTS Back to Search Results
Model Number 98430
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 24 sep 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).Device not returned.
 
Event or Problem Description
It was reported that the sutures broke upon insertion.No further information was provided.
 
Event or Problem Description
Per additional information received 26 sep 2018, there was no injury to the patient, and no medical intervention was required.
 
Additional Manufacturer Narrative
The device history record for lot aa7289r02 was reviewed and the product was produced according to product specifications.All information reasonably known as of 11 oct 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The used anchor sutures from the reported kit were returned for evaluation.Visual examination of the sutures confirmed them to be broken.The root cause could not be identified.All information reasonably known as of 13 nov 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 16FR DILATOR
Common Device Name
DH EF PERC PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key7909325
Report Number9611594-2018-00155
Device Sequence Number19297655
Product Code KGC
UDI-Device Identifier10680651984309
UDI-Public10680651984309
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K080253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2019
Device Model Number98430
Device Catalogue Number991098430
Device Lot NumberAA7289R02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer09/26/2018
10/31/2018
Initial Report FDA Received Date09/26/2018
Supplement Report FDA Received Date10/12/2018
11/14/2018
Patient Sequence Number1
Patient Age74 YR
-
-