| Catalog Number |
RFSPC-35-180 |
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product code: dqx.This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
The complainant reported that the roadrunner nimble hydrophilic wire guide would not come out of the holder, and it was split after it was finally removed.It was further noted that a section of the polymer material was missing from the distal end of the wire which resulted in exposure of the inner mandril wire.This product issue occurred prior to patient contact, and it did not result in any reported impact on the procedure.
|
| |
|