The customer obtained higher than expected vitros k+ and na+ results from a non-vitros biorad level 1 quality control fluid processed using vitros k+ and na+ slides in combination with a vitros 5600 integrated system.Biorad level 1, lot 47960 na+ results of 143.0, 142.7, and 158.1 mmol/l vs.The expected result of 115.0 mmol/l.Biorad level 1, lot 47960 k+ result of 3.70 mmol/l vs.The expected result of 2.50 mmol/l.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.The unexpected vitros na+ and k+ results were generated from non-patient fluids, however the investigation cannot conclude that patient sample results were not affected and would not be affected if the event were to recur undetected.There was no allegation of patient harm as a result of the event.This report is number one of two mdr¿s for this event.Two 3500a forms are being submitted for this event as two devices were involved.This report corresponds to ortho clinical diagnostics inc.Complaint numbers (b)(4).
|