| Model Number |
1884080EM |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the blade found that visually, the distal tip had broken off the inner cutter.The portion that became detached was not returned; however, it would have measured approximately 0.24¿ in length.The break occurred at the first proximal tooth valley and proceeded to the second.The outer assembly was deformed.There was no evidence of concentricity issues or bends in the returned tube / shaft assemblies.There was uneven wear at the cutting tips.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) via a manufacturer representative reported that during a transphenoidal surgery, the inner cutting cannula fractured off.This happened with two different blades using the same handpiece.After the second blade fractured, the handpiece was switched as the hcp thought the handpiece was causing the issue; however, a third blade fractured.All parts three pieces of the blades were removed from the patient using a pickup instrument with no issue.The power console was set at 5k and on oscillating movement.The procedure was delayed approximately 25 minutes due to switching the blades and handpiece.The case was completed using a different blade.No further actions was taken.There was no patient impact.
|
| |
|