| Catalog Number |
DTDW-35-110-10 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product code: dqx.This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
The complainant reported that the reuter disposable tip deflecting wire guide was not deflecting after the first use.The physician was attempting to reposition a stent in the renal artery.It was further noted that the wire guide was stretched from the solder joint to the distal tip.There were no reported serious injuries to the patient.
|
| |
|
Search Alerts/Recalls
|