One 60-5274-032 was received in opened original packaging, the reported catalog and lot numbers were verified.The device was visually inspected per drawings.The manufacturing process was reviewed regarding the origins of the defect.Visual inspection found that the sheath had become detached from the device and was able to slide along the molded tube.Inspection of the flanged end of the sheath found no evidence of damage or deformation.Possible causes of sheath detachment includes improper seating of the sheath during manufacture (misaligned or improper rotation).The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 24 complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: cautions and warnings: examine this device prior to use for damage.Do not use if damage is found.This issue will continue to be monitored through the complaint system to assure patient safety.
|