| Catalog Number |
2102-0410 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been one other event for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Reamer shaft handle broke under usage.Replacement handle was used.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding crack/fracture involving a cutting edge handle was reported.The event was confirmed.Conclusions: the investigation confirmed the reported event based on the supplier's analysis which indicated the reported event is confirmed and attributed to wear.
|
| |
|
Event or Problem Description
|
|
Reamer shaft handle broke under usage.Replacement handle was used.
|
| |
|