Catalog Number 103579
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that during set up with two units of polyflux 170h dialyzers, yellowish spots were observed between the fibers.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
One device was received for evaluation and one device was not returned; therefore, a device analysis could not be completed.A visual inspection was performed on the returned sample and it was noted that product was wet and yellowish coating was found on the fiber.A materials analysis could not detect ferrous.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 170H
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
hechingen
MDR Report Key7910632
Report Number9611369-2018-00113
Device Sequence Number1936647
Product Code KDI
UDI-Device Identifier07332414037932
UDI-Public(01)07332414037932
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K043342
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Catalogue Number103579
Device Lot Number7-4724-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/31/2018
Initial Report FDA Received Date09/26/2018
Supplement Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-